| Sr. No. |
Description |
Standard |
Optional |
|
1
|
M.O.C.
|
Glass body with SS-316 wetted parts.
|
Glass body with teflon coating on top bottom flanges &
Hastelloy-C internal wetted parts.
|
|
2
|
Maximum Working Pressure
|
6kg/cm2 (85psig) & 3Kg/cm2 for jacketted glass body.
|
N.A.
|
|
3
|
Maximum Working Temperature
|
150o C
|
N.A.
|
|
4
|
Stirrer RPM
|
50 - 1450 infinite variable
|
upto 3000
|
|
5
|
Heating System
|
Bottom plate heater
|
Jacketted Glass body with oil heating system.
|
|
6
|
Drive
|
DC
Motor & Drive
|
Flame
Proof AC motor & Invertor Drive
FLP
IIC Motor
|
|
7
|
Shaft Sealing
|
Stuffing
box gland type packing cones & 'O' rings
|
Zero
leakage maintenance free magnetic
drive coupling
|
|
8
|
Gasket & Bushes
|
Teflon with split clamp type quick opening system.
|
N.A.
|
|
9
|
Temp. Controller
|
S.S. control panel with Microprocessor
based P.I.D. Temperature Controller & alarm system.
|
a) Remotely controlled through PC /Panel
with monitoring, plotting & datalogging of parameters.
b) Flameproof Enclosure with
digital Temperature / Pressure / RPM indicators in
it, mounted on the autoclave trolley so that control
panel can be remotely mounted at a distance of upto
10m in safe non-flameproof area.
|
|
10
|
Stirrer
|
Pitch blade turbine
|
Hollow shaft with gas entrainment type impeller (specially
for hydrogenation) propeller, anchor, Helix.
|
|
11
|
Mounting
|
SS Trolley
|
SS table.
|
|
12
|
Standard Fittings on
autoclave
|
Pressure guage, vent valve, Pressure release valve, Internal
cooling coil, thermowell, gas inlet-sparger / liquid
sampler.
|
Ball valve for powder/liquid inlet, digital pressure indicator
with alarm & solenoid valve
|